Lag Time in AlternativeLand

Are Alternative Site Meters Dangerous?

By David Mendosa

Last Update: December 19, 2001

Blood glucose meters that let you test on places where you have few nerve endings are much less painful than traditional fingertip meters. These so-called alternative site meters are the best thing for people with diabetes since the first meters became available more than 30 years ago.

Two German scientists stirred up a hornet’s nest.

Or are they?

It seems that all is not perfect in alternativeland. In fact, they may be downright dangerous. Under certain circumstances they might not warn you quickly enough that your blood glucose level is going too low.

There are currently three different brands of alternative site meters on the market. First to be introduced, Amira Medical's AtLast became available in December 1999, but Roche Diagnostics bought the company in November 2001. Roche subsequently stopped marketing the AtLast meter, because Amira had sold only 20,000 to 30,000 units in the two years it was available.

However, following Amira's lead, LifeScan introduced the OneTouch Ultra, which takes only 5 seconds to give a reading. The TheraSense FreeStyle takes the least blood of all available meters. Another alternative site meter, the MediSense Sof-Tact, hides the test strip, lancet, and blood drop inside the meter.

There is a fourth alternative site meter. Beginning in mid-January 2001 Roche will market a test kit for its Accu-Chek Active meter including a different cap for its Softclix Lancet device, which will permit forearm testing. Roche studies found that alternative site testing measurements are equivalent to fingertip results if the patient's blood glucose is in a steady state, but the company does not recommend forearm testing within two hours after a meal or within two hours after injecting fast-acting insulin, according to Nancy Lonsinger, director of consumer marketing for Roche's U.S. market.

A hornet’s nest
Two German scientists, Karsten Jungheim and Theodor Koschinsky, stirred up a hornet’s nest when they presented their report on lags in alternative site tests at the ADA Scientific Sessions on June 25. They followed up with a letter published in the July 2001 issue of Diabetes Care.

Their preliminary studies indicate that tests with TheraSense's FreeStyle meter on the arm may lag about half an hour behind fingertip results. This could have serious consequences preventing timely detection of hypoglycemia if blood glucose levels were falling quickly.

They based their report on tests of just six people with type 1 diabetes on intensified insulin treatment. They did not rub the forearm skin before taking the blood glucose sample.

Note the rather unusual protocol of their test: After an overnight fast, they omitted the usual prebreafast insulin. Instead of breakfast, they each took 75 grams of glucose so that their blood glucose readings would be between 300 and 400 mg/dl. Then they got their usual (6 to 15 units) short-acting dose of insulin. The researchers watched their blood gluocose drop every 5 to 15 minutes until they reached either a steady state or hypoglycemia (below 60 mg/dl). Oral glucose compensated for the hypoglycemia. Some experts note that this large glucose dose followed by intravenous insulin is hardly a typical situation.

Rejoinder
A TheraSense rejoinder reported in the same issue of Diabetes Care showed little difference between fingertip and arm among 120 people the company studied—if they rubbed the arm first.

In these initial TheraSense studies using single points in time with many subjects, they saw close correlation between forearm, upper arm, back of the hand, thigh, and calf. But the correlation of all of these sites to the finger was not as good.

"This is in keeping with the circulatory physiology of the skin," according to e-mail from Mr. McGarraugh, TheraSense's director of chemistry. "The blood flow on the palms of the hands (including finger tips) and the soles of the feet is considerably faster than the blood flow in the skin of the arms, legs and abdomen. We believe the faster blood flow allows faster communication of the blood glucose changes. We currently are conducting the time course studies comparing thigh and finger. We will be reporting these results as soon as we have enough data to make sound conclusions."

Based on its studies to date, TheraSense has reached five conclusions, he says:

  1. Changes in glucose can be detected first on the finger, and the changes can lag on the forearm.

  2. Rubbing the test site prior to testing stimulates blood flow and minimizes the difference between the alternate site and the finger.

  3. The faster the changes the greater the difference between forearm and finger.

  4. There are significant differences from person to person. In some people the lag time is very short, and the differences between arm and finger are negligible. In other people the lag time can be significant—15-20 minutes.

  5. The lag is most significant in the detection of hypoglycemia. For some people the detection of hypoglycemia will be much more effective if the finger is used as the test site.

TheraSense continues to study the question, Mr. McGarraugh writes. "We are expanding to additional alternate sites and to specific patient populations (children, women with gestational diabetes, and people with peripheral vascular diseases). There will be an article in the Fall edition of Diabetes Technology and Therapeutics on this subject."

LifeScan has also studied the issue, although it has not yet published its conclusions. A LifeScan study of 42 people using their One Touch Ultra meter seemed to support Jungheim's and Koschinsky's finding. Finger tests tended to produce higher average blood glucose readings than forearm and thigh tests, especially an hour and an hour and one-half after meals. Not all patients, however, experienced this lag.

LifeScan is just now sending new recommendations to medical personnel. Alternative site testing, they say, should be done only two or more hours after the last meal, two or more hours after taking insulin, or two or more hours after exercise. Otherwise, LifeScan recommends fingertip testing.

Ron Ng, Ph.D., DABCC, FACB, director, medical and clinical affairs, Abbott Laboratories, MediSense Products, presented data on the accuracy of the Sof-Tact system for testing blood glucose in samples taken from the forearm. The Sof-Tact device increases perfusion by creating a vacuum over the sampling site. EGA of a clinical study of lay users demonstrated that the accuracy of the Sof-Tact system was clinically acceptable. Dr. Ng also described results of a meal tolerance test, an oral glucose tolerance test, and a hypoglycemia study designed to evaluate the accuracy of the device; in all three studies, EGA indicated that between 99.5% and 100% of the Sof-Tact results were clinically acceptable. Dr. Ng concluded by saying that the data support use of the device in both static and dynamic glucose conditions. Ron Ng, the director of medical and clinical affairs for MediSense products, told an FDA panel that clinical studies showed that at least 99.5% of the results of its Sof-Tact meter were clinically acceptable. He said that the data support the use of the Sof-Tact meter "in both static and dynamic glucose conditions." He apparently means when blood glucose is both stable and rapidly changing.

Meanwhile, anyone who uses any of the other alternative site meters and is concerned that his or her blood glucose is falling rapidly is well advised to rub or heat the test site or complement the alternative site test with a finger prick.

References:

"Arm and Thigh Capillary Blood Samples as Alternatives to Fingertip Sampling in Blood Glucose Monitoring." Unpublished LifeScan paper.

"Give your patients the freedom to choose where to test," LifeScan brochure.

Jungheim, Karsten, and Theodor Koschinsky, MD, "Risky Delay of Hypoglycemia Detection by Glucose Monitoring at the Arm," Diabetes Care, July 2001, Volume 24, pages 1303-1304. On-line at
http://care.diabetesjournals.org/cgi/content/full/24/7/1303-a Koschinsky, Theodor, and Karsten Jungheim, "Risky Delay of Hyper- and Hypoglycemia Detection by Glucose Monitoring at the Arm," Abstract #214-OR, 61st Scientific Sessions of the American Diabetes Association, June 25, 2001, Philadelphia, Pennsylvania. On-line at
http://38.204.37.95/am01/AnnualMeeting/Abstracts/NumberResults.asp?idAbs=214-OR

McGarraugh, Geoff, "Response to Jungheim and Koschinsky," Diabetes Care, July 2001, Volume 24, pages 1304-1306. On-line at
http://care.diabetesjournals.org/cgi/content/full/24/7/1304

McGarraugh, Geoff, Sherwyn Schwartz M.D., Richard Weinstein M.D., "Glucose Measurements Using Blood Extracted from the Forearm and the Finger," on-line at
http://www.therasense.com/img/TheraSensewhitepaper.pdf

Summary Minutes for the Meeting of the FDA's Clinical Chemistry and Clinical Toxicology Devices Panel on October 29, 2001, on-line at
http://www.fda.gov/ohrms/dockets/ac/01/minutes/3801m1.doc


This article originally appeared on Mendosa.com on August 1, 2001.


[Go Back] Go back to Home Page

[Go Back] Go back to Diabetes Directory

Never Miss An Update

Subscribe to my free newsletter “Diabetes Update”

I send out my newsletter on first of every month. It covers new articles and columns that I have written and important developments in diabetes generally that you may have missed.

Your Email Address

Most Popular Articles and Blog Posts


Advice For Newbies Diagnosis of Diabetes
Incorrect Terms Glycemic Values of Common American Foods
Glycemic Index The Normal A1C Level
Glycemic Values Controlling the Dawn Phenomenon
The Biggest Diabetes Scams The Food Insulin Index Trumps Carb Counting
David’s Guide to Getting our A1C Under 6.0 Chia Seeds
What Really Satisfies Snake Oil Supplements

diabetes supplement
Never Miss An Update!